Statistics for PT

z-scores in proficiency testing and EQA: formula, bands and meaning

A z-score turns the difference between a laboratory’s result and the assigned value into a number that reads the same for any measurand. This guide shows how it is calculated, what satisfactory, questionable and unsatisfactory mean, how often chance alone produces a flag, and how to check a score in a report.

Byline
By the EqasOnline team
Last reviewed
Reading time
8 min read
On this page
  1. The z-score formula
  2. What the sign tells you
  3. Satisfactory, questionable, unsatisfactory
  4. Why the limits sit at 2 and 3
  5. How often chance alone flags a result
  6. When σpt does not match the real spread
  7. Checking a score in a report
  8. Related scores: z′, SDI and percent difference
  9. A score is a signal, not a verdict

The z-score formula

A z-score compares a participant’s result with the assigned value, measured in units of the standard deviation for proficiency assessment:

z = (x − xpt) / σpt

The three quantities in a z-score.
SymbolMeaningWhere it comes from
xThe participant’s resultWhat the laboratory reported
xptThe assigned valueSet by the provider: from how the item was made, from a reference value or from the consensus of participants. See how the assigned value is set.
σptThe standard deviation for proficiency assessmentSet by the provider, ideally from what the results are needed for. See choosing σpt.
The three quantities in a z-score.

The top line is the error in the units of the measurement. Dividing by σpt turns it into a pure number, so a z-score for glucose, lead or moisture reads on the same scale.

Example: lead in drinking water

x = 10.9 µg/L, xpt = 10.0 µg/L and σpt = 0.75 µg/L. So z = (10.9 − 10.0) / 0.75 = 0.9 / 0.75 = +1.2. The result is 1.2 σpt above the assigned value.

What the sign tells you

A positive z means the result was high and a negative z means it was low. The size says how far. One high score tells you little, but a run of positive scores across rounds points to a consistent positive bias, and the direction narrows down the causes worth looking for.

Warning: subtracting the wrong way round

Calculating (xpt − x) instead of (x − xpt) flips the sign. The band is unchanged, but an investigation would start looking for a low bias when the result was high.

Satisfactory, questionable, unsatisfactory

The conventional z-score bands described in ISO 13528.
Size of zClassificationWhat it usually means
2.0 or lessSatisfactoryNo action beyond routine review
Above 2.0 and below 3.0QuestionableA warning signal: review it and watch for repeats
3.0 or moreUnsatisfactoryAn action signal: investigate it as a nonconformity
The conventional z-score bands described in ISO 13528.

These are the conventional limits, and most schemes use them. A scheme may set others, and its scheme plan says which apply. Two boundary cases are worth knowing. A score of exactly 2.0 is satisfactory, because the questionable band starts above 2.0. A score of exactly 3.0 is unsatisfactory, because that band starts at 3.0.

Warning: rounding before you classify

Classify on the score the provider reports, under the rounding rule in its scheme plan. If you recalculate a score yourself, keep full precision and round only at the end. An unrounded 2.04 is reported as +2.0 by a scheme that gives one decimal place and as +2.04, questionable, by a scheme that gives two.

Example: three results in one round

Moisture in wheat flour, with xpt = 13.20 g/100 g and σpt = 0.30 g/100 g.

Three laboratories scored against the same assigned value. Example data.
LaboratoryResultx − xptzClassification
A13.62+0.42+1.4Satisfactory
B12.54−0.66−2.2Questionable
C14.10+0.90+3.0Unsatisfactory
Three laboratories scored against the same assigned value. Example data.

Laboratory C landed exactly on the boundary, and 3.0 is unsatisfactory.

Why the limits sit at 2 and 3

The bands rest on a simple model. Suppose the results of competent laboratories scatter around xpt in a normal distribution whose standard deviation equals σpt. Their z-scores then follow the standard normal curve: centred on 0, with a standard deviation of 1.

Share of z-scores in each range when σ_{pt} equals the real spread of competent laboratories.
Range of zShare of scoresRoughly
Within ±168.3 %2 in 3
Within ±295.4 %19 in 20
Between 2 and 3, either side4.3 %1 in 23
At or beyond ±30.27 %3 in 1,000
Share of z-scores in each range when σpt equals the real spread of competent laboratories.

So the limits are not arbitrary. A score beyond ±2 is uncommon for a competent laboratory, which makes it a sensible warning. A score beyond ±3 is rare enough that chance becomes an unlikely explanation, which is why it calls for action. The exact 95 % points of the normal curve are ±1.96; PT uses the round numbers 2 and 3.

How often chance alone flags a result

A laboratory with no problem at all gets a questionable score about 4.3 % of the time, or roughly 1 result in 20. With many measurands, those small chances add up.

Example: a water laboratory’s year

Twelve parameters in each of four rounds gives 48 scores a year.

  • Questionable by chance: 48 × 0.043 ≈ 2 a year.
  • Unsatisfactory by chance: 48 × 0.0027 ≈ 0.13 a year, or about one every 7 to 8 years.
  • A round with all 12 scores satisfactory: 0.95412 ≈ 0.57. Only about 57 % of rounds come back clean, even when nothing is wrong.

Two consequences follow.

  • One questionable score is not evidence of a fault. Record it, review it and watch the next rounds.
  • Patterns are evidence. Two questionable scores in a row on the same measurand happen by chance only about once in 500 pairs of rounds, and an unsatisfactory score is rare enough to be treated as a nonconformity.

Warning: real data have heavier tails

Real rounds include blunders, wrong units and transcription errors, so extreme scores turn up more often than the normal curve predicts. That is one reason providers use robust statistics such as Algorithm A. The percentages describe what chance alone does. They are not a promise about any one round.

When σpt does not match the real spread

The percentages hold only when σpt equals the spread that competent laboratories really achieve. Providers often set σpt from what the results are needed for instead, and then the two can differ. If the real spread is twice σpt, a competent laboratory’s z-scores have a standard deviation of 2, not 1.

How the share of flagged results changes with the choice of σ_{pt}.
σpt compared with the real spreadCompetent laboratories beyond ±2Beyond ±3
Twice as wide0.006 %Almost none
Equal4.6 %0.27 %
Two-thirds as wide18 %4.6 %
Half as wide32 %13 %
How the share of flagged results changes with the choice of σpt.

A σpt wider than the real spread hides problems

Example: zinc in wastewater

Competent laboratories agree with a standard deviation of 0.12 mg/L, but the scheme uses σpt = 0.30 mg/L. A laboratory with a calibration bias of 0.36 mg/L, three times the real spread, scores z = 0.36 / 0.30 = +1.2: satisfactory. Measured against the real spread, the same error would be 0.36 / 0.12 = 3.0.

That is acceptable if 0.30 mg/L really is what the users of the data need: the score then says the result is good enough for its use. It does not say the method agrees with its peers, so look at where the scores sit, not only at their band.

A σpt narrower than the real spread flags good laboratories

If σpt is half the real spread, about a third of competent laboratories fall outside ±2 in every round, and about 13 % beyond ±3. The chance rates above no longer apply, and a questionable score says little. If the narrow σpt reflects a real requirement that current methods cannot meet, the flags are honest. If it is simply a poor choice, they are noise.

A check you can do

When σpt matches the real spread, the robust standard deviation of all participants’ z-scores in a round is close to 1. Well above 1 means σpt is tighter than laboratories achieve; well below 1 means it is wider. If the report shows every participant’s score, or a histogram of them, you can judge this by eye.

Checking a score in a report

If a report gives your z, xpt and σpt, you can recover the result that was scored:

x = xpt + z × σpt

This catches transcription errors. Suppose a potassium report shows z = −2.8 with xpt = 4.20 mmol/L and σpt = 0.10 mmol/L. The scored result was 4.20 − 2.8 × 0.10 = 3.92 mmol/L. If the instrument printout says 4.29, two digits were swapped when the result was entered, and the measurement itself would have scored z = +0.9.

To score a whole round yourself, the free z-score calculator takes a list of results, finds the robust mean and standard deviation with Algorithm A and shows every z-score with its band. How to read a PT report gives the full checklist.

  • z′ (z-prime) adds the uncertainty of the assigned value to the denominator. Providers use it when that uncertainty is more than 0.3 σpt, so that participants are not penalised for an assigned value that is not known well enough. It is read on the same bands as z. See z′, ζ and En scores.
  • SDI (standard deviation index), common in clinical EQA reports, is the same calculation with the peer group’s mean and standard deviation in place of xpt and σpt. Because that spread comes from the round’s own results, it places a laboratory among its peers. It does not say whether the results are good enough for their use.
  • Percent difference, D% = 100 × (x − xpt) / xpt, leaves σpt out and is judged against a fixed limit set by the scheme.

A score is a signal, not a verdict

Every z-score depends on two choices the provider made: the assigned value and σpt. The same result can be satisfactory under one σpt and questionable under another, so read the scheme plan before reading too much into one number. Then size the response to the signal.

Take the free lesson

Module 4 · Performance scores

Five lessons and a practice case: the z score and its bands with an interactive simulator, what the bands mean statistically, scores that bring uncertainties in, percent difference and qualitative results, and choosing a score. The course is free in the EqasOnline Academy; lessons open after you create a free account.

Take the free lesson

Scoring PT rounds?

In EqasOnline one server-side engine scores every round the same way, following ISO 13528 methods. Every result gets a z-score and a % bias, each band is written in words, and z′ is shown beside z in the provider’s review matrix and the parameter-wise report. Labs see their z-scores, peer groups and trends across rounds on their dashboard.