Reports and follow-up
PT trends and Youden plots: reading scores across rounds and samples
One PT score is a snapshot: you cannot tell whether it landed where it did by chance. Across rounds, and across the samples in one round, chance tends to cancel out and a small, steady problem shows itself long before it produces an unsatisfactory score. This guide covers the trend chart, the Youden plot and results at several levels.
- Byline
- By the EqasOnline team
- Last reviewed
- Reading time
- 7 min read
On this page
Why look across rounds
ISO/IEC 17025:2017 (7.7.3) asks laboratories to analyse monitoring data, PT included, and act when it falls outside criteria defined in advance. ISO 15189:2022 (7.3.7.3) asks medical laboratories to review EQA results and act on them. ISO 13528:2022 (clause 10) describes graphical methods, including plots of scores across rounds and the Youden plot.
Build the chart
- One chart per measurand, and one per level if the scheme sends more than one.
- Your score on the vertical axis, with lines at ±2 and ±3, and the rounds in date order along the bottom. Use one kind of score throughout.
- Mark events in your laboratory: reagent and calibrator lots, recalibration, service or repair, a new instrument or method, staff changes.
- Mark scheme changes too. A new way of setting the assigned value or σpt can move every score while your laboratory stays the same. Where σpt varies between rounds, keep the difference in units alongside z.
Five patterns
| Pattern | What it looks like | Typical causes |
|---|---|---|
| Drift | Scores climb or fall steadily over several rounds | Ageing reagent or calibrator, a weakening lamp or electrode, a stock standard slowly evaporating |
| Step change | A jump between two rounds that then stays | A new reagent or calibrator lot, recalibration, a new instrument, method or standard |
| Run on one side | Many scores in a row on the same side of zero, not necessarily large | A small, consistent bias |
| Repeated warnings | Questionable scores more often than about 1 in 20, especially of one sign | A bias close to the limit, or poor precision |
| Widening scatter | Scores swing further either side of zero | Lost precision: a manual step, a worn pipette, new or rushed staff, an unstable instrument |
What chance looks like
For a laboratory with no bias, in a scheme whose σpt matches the spread of good laboratories:
- about two scores in three fall within ±1, and about 1 in 20 outside ±2;
- the sign is a coin toss, so six in a row on the same side of zero happens about once in 32 runs of six;
- two questionable scores in a row on the same measurand happen only about once in 500 pairs of rounds.
These are aids to judgement, not rules. Some laboratories borrow control-chart rules, such as a set number of consecutive scores on one side of zero, as triggers in their own procedure. That is a local choice written down in advance, not a requirement of ISO 13528 or the accreditation standards. z-scores in proficiency testing explains where the chance figures come from.
Satisfactory, but not fine
Example: sodium in bread
σpt is fixed at 6 % of the assigned value. Six rounds give z-scores of 1.4, 1.7, 1.5, 1.9, 1.6 and 1.8, every one of them positive.
Every score is satisfactory. But six positive scores in a row is about a 1 in 64 event for a laboratory with no bias, and the scores lie within 0.5 of each other, so precision is good. Their mean is 9.9 / 6 = 1.65: a stable positive bias of about 1.65 σpt, or 1.65 × 6 % ≈ 10 % high.
A 10 % bias can matter for a salt claim on a label, whatever the scheme’s bands say. Look for the cause now (calibration standard, recovery, blank) rather than waiting for a questionable score. Size a bias in units, z × σpt, and judge it against your customers’ needs, which may be tighter than the scheme’s σpt.
Several measurands at once
When several measurands move together, ask what they share: an instrument, a calibration standard, a preparation step, an analyst. Suppose copper, zinc and lead in water all drop from about 0 to about −1.3 after a new mixed standard was made, and stay there, while mercury, measured by another technique, does not move. One cause explains all three: a standard stronger than its label, perhaps through evaporation, makes samples read low.
If every participant moved in one round, suspect the item or the assigned value, and check the round’s statistics. A shift that persists across several rounds, with different items each time, is not an item problem.
The Youden plot: one point per laboratory
When a round contains two similar items, A and B, each laboratory becomes one point: its z for A along the bottom, its z for B up the side, on the same scale. Many reports draw the ±2 square and the diagonal where the two scores are equal. The plot reads best when A and B are close in level.
| Where the point sits | What it suggests | Where to look |
|---|---|---|
| Near the centre | Nothing to see | Nowhere |
| Near the diagonal, away from the centre | Systematic error: both items moved the same way | Calibration, standard values, blank or recovery |
| Far from the diagonal | Something affected one item, or poor repeatability | A mix-up, contamination, a dilution or entry slip; precision if it recurs |
Two numbers split a point. The average of the two scores measures what the items share. Their difference measures what differs between them.
Example: two laboratories, one round
Deoxynivalenol in wheat flour. Laboratory 7 scores +2.6 on item A and +2.2 on item B: average +2.4, difference 0.4, so mostly systematic. Start with calibration. Laboratory 12 scores +2.8 on A and −0.6 on B: average +1.1, difference 3.4, so item A alone. Start with how A was handled and entered.
Look at the whole cloud too. In many schemes most points form a band along the diagonal: each laboratory carries its own bias into both items, so systematic differences dominate the spread. A round cloud means random error dominates. One method group clustered along the diagonal points to a method bias.
Warning: inside the square is not always fine
A point with +1.6 on one item and +1.7 on the other passes both scores, but it sits on the diagonal. Round after round, that is a consistent bias: the run on one side of zero described above.
To see the patterns for yourself, the free Youden plot in the EqasOnline Academy plots pairs of scores as you enter them.
Results at several levels
When a scheme sends several levels, compare your differences in units and in percent. Take copper in water, with σpt at 10 % of the assigned value at every level.
| Laboratory | At 50 µg/L | At 200 µg/L | At 800 µg/L | Pattern |
|---|---|---|---|---|
| K | +12 µg/L (+24 %, z = +2.4) | +12 µg/L (+6 %, z = +0.6) | +12 µg/L (+1.5 %, z = +0.15) | Constant |
| L | +4 µg/L (+8 %, z = +0.8) | +16 µg/L (+8 %, z = +0.8) | +64 µg/L (+8 %, z = +0.8) | Proportional |
| M | +0.5 µg/L (+1 %, z = +0.1) | +1 µg/L (+0.5 %, z = +0.05) | −144 µg/L (−18 %, z = −1.8) | Top of the range |
- Constant error adds the same amount at every level, so its percentage shrinks as the level rises. With σpt as a percentage, the largest z is at the lowest level. Look at the blank, the baseline, contamination or an interference that adds a fixed amount.
- Proportional error is the same percentage at every level, so the difference in units grows with concentration. With σpt as a percentage, z is the same at every level. Look at calibrator or standard values, dilution factors and recovery.
- A range problem shows at one end only. At the top: non-linearity, detector saturation, or results beyond the calibrated range without dilution. At the bottom: the blank, or working too close to the limit of quantification. Laboratory M’s −1.8 is satisfactory, but 18 % low is a lot.
Across levels and rounds you can fit a straight line, result = a + b × xpt. The intercept a estimates the constant error and b − 1 the proportional error. Results of 46.0 at an assigned value of 40 and 214.0 at 200 give b = 168 / 160 = 1.05 and a = 46.0 − 1.05 × 40 = 4.0: a 5 % proportional error and a constant error of 4.0 units. Two points always fit exactly, so confirm with more.
Deciding what to do
- A single questionable score with no pattern: review it, record it and watch.
- A pattern of any of the five kinds, or a shared shift: investigate it as a potential nonconformity, even when every score is satisfactory. See investigating an unsatisfactory PT result.
- Write your trend criteria into your procedure in advance, so the decision does not depend on who reads the chart.
For the report each score comes from, and the peer-group statistics beside it, see how to read a PT report.
Take the free lesson
Module 5 · Reading reports and trends
Four lessons and a practice case: reading a PT report, peer groups and method bias, trends across rounds, and Youden plots with an interactive tool. The case follows three years of creatinine results through a reagent change. The course is free in the EqasOnline Academy; lessons open after you create a free account.