Statistics for PT

The assigned value in proficiency testing: five ways to set it

Every z-score is measured from the assigned value, so an error in it becomes an error in every participant’s score at once. This guide sets out the five ways ISO 13528 gives for setting it, what each is traceable to, the four situations in which a consensus of participants misleads, and how to judge whether the value is known well enough to score against.

Byline
By the EqasOnline team
Last reviewed
Reading time
9 min read
On this page
  1. What the assigned value is
  2. Five ways to set the assigned value
  3. What a good score means depends on the value behind it
  4. Four ways a consensus misleads
  5. Checking a consensus against a reference value
  6. The uncertainty of the assigned value
  7. Method-group assigned values
  8. Choosing, and writing it down

What the assigned value is

The assigned value, xpt, is the provider’s best estimate of the true value of the measurand in the PT item. Every z-score is measured from it. The provider also needs its standard uncertainty, u(xpt), and, where possible, a statement of what the value is traceable to.

ISO 13528:2022, clause 7, describes five ways to set it. The provider chooses one when it plans the scheme and writes it into the scheme plan. It is not picked after the results arrive.

Five ways to set the assigned value

They run from the most independent of the participants to the least.

1. Formulation

The provider makes the item from known amounts: a pure substance weighed into a base material that contains none of the measurand, or a measured amount of it. xpt is calculated from the masses or volumes. It fits simple, well-defined matrices such as spiked waters, synthetic solutions and gas mixtures. Its weak points are analyte already in the base material, added analyte that does not behave like the native form, losses during preparation and storage, and poor mixing in solids.

Example: nickel in water, made up by mass

Base water: 1.9800 kg, measured at 0.50 µg/kg nickel, so it carries 0.99 µg. Added: 20.00 g of a 2.000 mg/kg nickel standard, which carries 40.00 µg. Together that is 40.99 µg of nickel in 2.0000 kg, so xpt = 20.5 µg/kg. Ignoring the base water would give 20.0 µg/kg, 2.4 % low.

2. A certified reference material

The PT item is a certified reference material, or is made from one, and xpt is the certified value. It fits when such a material exists in a matching matrix and at a useful level, and the scheme can afford one unit per participant. The weak points are cost, the few matrices and levels available, a published value that participants may recognise, and a material that some methods read differently from routine samples.

3. A result from one laboratory

One competent laboratory measures the items, usually with a reference method or against a closely matched reference material, and its result becomes xpt. It fits comparisons that circulate an artefact, clinical analytes with a reference measurement procedure, and small specialist schemes. Everything rests on one laboratory, with no independent check of its bias.

4. Consensus of expert laboratories

A selected group of laboratories with demonstrated competence measure the items, and xpt is a robust mean of their results. It fits when no reference material exists, the stakes are high, and a few laboratories are clearly better placed than the average participant. Experts can share a method bias, few results mean a larger uncertainty, and the selection must be justified and recorded.

5. Consensus of participants

xpt is a robust mean of the participants’ own results, usually from Algorithm A. It is the most common choice: it fits when there are many participants, no reference is available, or the measurand is defined by the method. It has no traceability of its own, because it describes what the participants measure, and it exists only after the round closes.

The five options compared.
OptionIndependent of participantsu(xpt) fromTraceable toWatch for
FormulationYesWeighing, volumes, purityThe SI, via balances and purityNative analyte, losses, recovery
Certified reference materialYesThe certificate, U ÷ kWhat the certificate statesCost, matrix match, recognition
One laboratoryYesThat laboratory’s uncertainty budgetIts calibration chainOne laboratory’s bias
Expert consensusPartlyThe experts’ spread and numberThe experts’ methodsShared method bias, few results
Participant consensusNo1.25 × s* / √pNothing of its ownShared bias, two populations, few results
The five options compared.

Warning: independent is not the same as right

A formulated value that ignores the native analyte, or a reference laboratory with an unnoticed bias, is wrong for every participant at once. Independence removes the participants’ influence; it does not remove the need to check the value.

What a good score means depends on the value behind it

A result is traceable when it can be related to a stated reference, usually an SI unit, through a documented, unbroken chain of calibrations. A formulated value, a certified value and a reference laboratory’s value can all be traceable. A consensus is the centre of what participants measure, and is traceable to nothing in particular.

This changes what a good score means. Against a traceable xpt, a z near zero says the result is close to the true value. Against a consensus, it says the result is close to its peers. Usually those are the same thing. The four cases below are when they are not.

Four ways a consensus misleads

1. A bias most participants share

If most participants use one method, calibrator or reagent that reads high or low, the consensus moves with them. Their z-scores look fine, and the few laboratories that are right are scored against a biased value.

Example: vitamin D in serum

Most participants use immunoassay kits that, in this invented round, read low in this material; a few use LC-MS/MS. The consensus is x* = 55.9 nmol/L, a reference laboratory finds 61.5 nmol/L, and σpt = 5.0 nmol/L. A result of 49.5 nmol/L scores z = −1.3 against the consensus, satisfactory, and −2.4 against the reference value, questionable.

2. A material that is not commutable

PT items are often stabilised, freeze-dried, spiked or pooled. If some methods respond to the item differently from the way they respond to routine samples, the item is not commutable for those methods. Their results differ for a reason that never affects a patient or customer sample, so one all-method xpt is unfair to them, and a reference value is no fairer.

3. Two method populations

When two methods measure slightly different things, the results form two clusters. A robust mean does not choose between them: it lands in the gap, where nobody reported, and the robust standard deviation swells to several times the spread inside either cluster. Plot the results, as a dot plot or histogram, before trusting any consensus.

4. Too few results

With few results, one laboratory can move the robust mean, the robust standard deviation is poorly estimated, and u(xpt) is large. Robust statistics need numbers to work.

Checking a consensus against a reference value

ISO 13528 clause 7 covers comparing a consensus with an independent reference value when one is available. A practical test is to compare the difference with twice the combined standard uncertainty of the two values.

Example: the vitamin D round, checked

Consensus: x* = 55.9 nmol/L from 17 results with s* = 4.33 nmol/L, so u = 1.25 × 4.33 / √17 = 1.31 nmol/L. Reference: 61.5 nmol/L with u = 1.0 nmol/L. The difference is 5.6 nmol/L. Twice the combined uncertainty is 2 × √(1.31² + 1.0²) = 2 × 1.65 = 3.3 nmol/L. Since 5.6 is larger than 3.3, the consensus is significantly low. Look for a shared bias or a problem with the material before using it.

The uncertainty of the assigned value

The assigned value is an estimate, so it has a standard uncertainty. ISO 13528 builds it from up to four parts: characterising the value, differences between items, and changes in transit and over the round.

u(xpt) = √(uchar2 + uhom2 + utrans2 + ustab2)

If the items pass their homogeneity and stability checks with room to spare, providers often treat the last three as negligible. If an item only just passes, they belong in the total. The first part depends on how the value was set.

  • A participant consensus: u(xpt) = 1.25 × s* / √p, for a robust mean of p results with robust standard deviation s*. The factor 1.25 allows for a robust estimate being less precise than a plain mean of normal data. Because of the √p, it takes four times as many results to halve the uncertainty.
  • A value from a certificate or a reference laboratory: a certificate gives an expanded uncertainty U and its coverage factor k, and the standard uncertainty is u = U / k. A certificate reading 0.284 ± 0.012 mg/kg with k = 2 gives u = 0.006 mg/kg. Copying the ± figure as if it were u doubles the uncertainty for no reason.

Is it small enough to ignore?

ISO 13528 treats the uncertainty of the assigned value as negligible when it is no more than 0.3 σpt. At that limit, including it would change scores by less than 5 %.

u(xpt) ≤ 0.3 σpt

When the rule fails, ISO 13528 points the provider to three responses.

  1. Find a better assigned value: more results, a reference laboratory or a reference material with a certificate.
  2. Score with z′, which adds the uncertainty to the denominator: z′ = (x − xpt) / √(σpt2 + u2(xpt)). It is judged on the same bands as z.
  3. Keep z and say so: tell participants that the uncertainty is not negligible, so scores near the limits need care.

Example: mercury in sediment, where the rule fails

From 16 results with s* = 0.064 mg/kg, u(xpt) = 1.25 × 0.064 / 4 = 0.020 mg/kg. σpt = 0.050 mg/kg, so 0.3 σpt = 0.015 mg/kg and the rule fails. With xpt = 0.420 mg/kg, the z′ denominator is √(0.050² + 0.020²) = 0.0539 mg/kg.

Four results scored with z and with z′. Example data.
LaboratoryResult (mg/kg)zz′
P0.480+1.20+1.11
Q0.525+2.10, questionable+1.95, satisfactory
R0.575+3.10, unsatisfactory+2.88, questionable
S0.330−1.80−1.67
Four results scored with z and with z′. Example data.

z′ is always a little smaller in size than z, so it can only move a result towards satisfactory. That protects participants from the provider’s uncertainty, but it also loosens the criterion, so a better assigned value is the better fix. z′, ζ and En scores compares the scores that use uncertainty.

Method-group assigned values

When an item is not commutable, or methods define the measurand differently, a provider may set xpt separately for each method group. That is fair to each method, but it has limits.

  • A good z then means agreement with the group, not with the true value. A bias the whole group shares stays invisible.
  • Small groups give unstable values with large uncertainties.
  • The score depends on each participant reporting the right method code.

Choosing, and writing it down

Weigh four things: how independent and traceable the value needs to be, which reference materials and laboratories exist, how many participants there are, and cost. Where a provider has both a reference value and a consensus, comparing the two is a strong check on each. Record the choice and the reason in the scheme plan, and state in the report how the value was set, from how many results, and its uncertainty.

The other half of every score is the yardstick it is divided by: see choosing σpt. For the plan that holds both choices, see running a proficiency testing scheme.

Take the free lesson

Module 3 · Assigned value and σpt

Five lessons and a practice case: the ways to set an assigned value, when a consensus misleads, Algorithm A with an interactive calculator, choosing σpt, and the uncertainty of the assigned value. The course is free in the EqasOnline Academy; lessons open after you create a free account.

Take the free lesson

Scoring against a consensus?

In EqasOnline one server-side engine finds the consensus the same way every round, following ISO 13528 methods: Algorithm A for 12 or more results, median and MADe for 3 to 11, and no z-score from fewer than three results. The uncertainty of the assigned value is checked against 0.3 σpt, and a detailed evaluation report prints every Algorithm A iteration for your assessor.