σpt, the standard deviation for proficiency assessment, is the yardstick behind every z-score. It decides how far from the assigned value a result may be before a scheme calls it questionable or unsatisfactory. This guide sets out the five approaches ISO 13528 describes, works an example of each calculation, and shows what each choice does to the scores.
- Byline
- By the EqasOnline team
- Last reviewed
- Reading time
- 6 min read
On this page
What σpt decides
σpt is the denominator of the z-score, z = (x − xpt) / σpt. It decides how far from the assigned value a result may be before the scheme calls it questionable (beyond 2 σpt) or unsatisfactory (3 σpt or more). It is a statement of the performance the scheme expects, not a property of the test item. Halve σpt and every z doubles.
The same yardstick is used around the score as well: the homogeneity and stability of the items, and the uncertainty of the assigned value, are each judged against 0.3 σpt. The provider chooses the approach when it plans the scheme and writes it into the scheme plan. ISO 13528:2022, clause 8, describes five.
Five ways to set σpt
1. Fitness for purpose: a prescribed value
σpt is set from what the results are used for: a regulatory requirement, a clinical decision, a customer specification, or the judgement of experts in the field. It can be a fixed amount, a percentage of xpt, or a combination such as the larger of 0.5 µg/L and 10 %, which stops σpt becoming impossibly tight near zero. Clinical schemes often derive it from biological variation; environmental schemes from the tolerance a regulation allows.
- For: the score means the same thing every round, and it answers the question participants care about: is my result good enough for its use?
- Against: it needs a defensible requirement. If it is far tighter than current methods can achieve, competent laboratories fail.
2. Previous rounds
σpt comes from the spread laboratories achieved before, for example a typical robust standard deviation from earlier rounds at similar levels, or a fitted line of robust standard deviation against concentration.
- For: realistic and stable.
- Against: it freezes past performance. If methods improve, the yardstick stays loose; if past rounds were poor, it was loose from the start.
3. A general model
The best known is the Horwitz equation from food analysis, with the concentration c and σ both as mass fractions:
σ = 0.02 × c0.8495
Thompson’s modification replaces it at the extremes: below a mass fraction of 1.2 × 10⁻⁷ (120 µg/kg) use a relative standard deviation of 22 %, and above 0.138 use σ = 0.01 × c0.5.
Example: copper in animal feed at 10 mg/kg
c = 10 mg/kg = 1 × 10⁻⁵. σ = 0.02 × (10⁻⁵)0.8495 = 1.13 × 10⁻⁶, which is 1.13 mg/kg, a relative standard deviation of 11.3 %. At 50 µg/kg the original equation gives 25 %, so Thompson’s 22 % applies instead: σ = 11 µg/kg.
- For: available before any results arrive, at any level.
- Against: it describes the reproducibility typical of past collaborative studies, not what the results are needed for, and it may not suit a particular method or matrix.
4. A precision experiment
If a collaborative study, for example one run to ISO 5725-2, has given the method’s repeatability SD sr and reproducibility SD sR, and each laboratory reports the mean of m replicates:
σpt = √(sR2 − sr2 × (1 − 1/m))
Averaging m replicates removes part of the within-laboratory scatter, which is why part of sr2 is taken away. With m = 1, σpt = sR.
Example: chemical oxygen demand in wastewater
sr = 4.0 mg/L, sR = 10.0 mg/L and m = 3. σpt = √(10.0² − 4.0² × (1 − 1/3)) = √(100 − 10.7) = √89.3 = 9.45 mg/L.
- For: grounded in measured method performance.
- Against: it holds only for the method, matrix and level range studied, and only when participants use that method. Like the general model, it says what the method can do, not what users need.
5. This round’s data
σpt is the robust standard deviation, s*, of the participants’ results in the same round, for example from Algorithm A.
- For: needs no outside information.
- Against: the yardstick moves with the participants, so scores cannot be compared between rounds, and small rounds give an unstable s*. Worst of all, it fixes the share of satisfactory scores in advance, as the next two sections show.
The same results, three choices
Example: iron in drinking water
xpt = 200 µg/L. Three candidate values: σpt = 15.0 µg/L (a prescribed 7.5 %), 18.0 µg/L (previous rounds) and 30.0 µg/L (this round’s s*, in a poor round).
| Laboratory | Result (µg/L) | z with 15.0 | z with 18.0 | z with 30.0 |
|---|---|---|---|---|
| A | 214 | +0.93 | +0.78 | +0.47 |
| B | 236 | +2.40, questionable | +2.00, satisfactory | +1.20 |
| C | 247 | +3.13, unsatisfactory | +2.61, questionable | +1.57 |
| D | 168 | −2.13, questionable | −1.78 | −1.07 |
With the prescribed value, three of the four need attention. With this round’s spread, all four look satisfactory. The results did not change; only the yardstick did.
Why this round’s own SD makes about 95 % look satisfactory
If σpt is the robust standard deviation of the same results, the z-scores have a robust standard deviation of about 1 by construction. For roughly normal data about 95 % then fall within ±2 and about 0.3 % beyond ±3, whether the round was excellent or poor. A reagent problem that doubles everyone’s scatter doubles σpt too, and the scores do not move. Real improvement never shows either.
Key point: what such a score can and cannot say
A score against the round’s own spread places a laboratory among this round’s participants. It cannot show how good the laboratories are, and it cannot be compared with the score from another round.
Choosing well
Prefer a value fixed in advance from what the results are needed for, wherever a defensible requirement exists, and check it against what laboratories achieve in practice, from previous rounds or a precision study, so the scheme is demanding but fair.
- Watch the gap. ISO 13528 clause 8 suggests comparing each round’s robust standard deviation with σpt over time. A persistent gap means the methods, or σpt, need attention.
- Keep it workable near zero. A percentage alone becomes impossibly tight at low levels; the larger of an absolute and a relative value avoids that.
- Fix it in the scheme plan. Whatever the choice, it is written down before the round and not changed after seeing the results. See running a proficiency testing scheme.
- State it in the report. Participants need σpt and how it was set to read their scores. See how to read a PT report.
σpt is one half of every score. The other half is the assigned value: how the assigned value is set covers the five options and the uncertainty that comes with each.
Take the free lesson
Module 3 · Assigned value and σpt
Five lessons and a practice case: the ways to set an assigned value, when a consensus misleads, Algorithm A with an interactive calculator, choosing σpt, and the uncertainty of the assigned value. The course is free in the EqasOnline Academy; lessons open after you create a free account.
Take the free lessonSetting up a scheme’s scoring?
EqasOnline scores every round with one server-side engine that follows ISO 13528 methods: Algorithm A for 12 or more results, median and MADe for 3 to 11, and no z-score from fewer than three results. The scoring and reports are already built in; your σpt and assigned-value choices stay yours.