Proficiency testing basics
External quality assessment (EQA): a guide for medical laboratories
External quality assessment (EQA) is how a medical laboratory checks its results against a value from outside its own walls. An EQA provider sends the same samples to many laboratories, each tests them like patient samples, and the provider reports how every result compares. This guide explains how a scheme runs, how results are judged, what ISO 15189 asks, and how to choose a scheme and use its reports.
- Byline
- By the EqasOnline team
- Last reviewed
- Reading time
- 8 min read
On this page
What external quality assessment is
In an EQA scheme, an independent provider sends samples of the same material to every participating laboratory. Each laboratory tests them and submits its results by a closing date. The provider compares every result with a target, called the assigned value, and sends each laboratory a report showing how it performed, identified only by a code.
EQA is the name medical laboratories use, and the one ISO 15189 uses. Outside medicine the same activity is called proficiency testing (PT), and ISO/IEC 17043, the standard for providers, uses that term. An EQA programme is proficiency testing, often with education added: case commentaries, method reviews and advice on interpretation. Proficiency testing: the complete guide covers the wider picture.
Key point: the value comes from outside
Everything else in a laboratory’s quality system compares the laboratory with itself. EQA is the regular check that brings in a value the laboratory did not set, on a sample it did not prepare.
EQA and internal quality control do different jobs
| Internal quality control | External quality assessment | |
|---|---|---|
| Question | Can the results of this run be released? | Do our results agree with an outside value and with our peers? |
| Material | Control material the laboratory buys or prepares | Samples sent by the provider, value unknown to the laboratory |
| Target | Set by the laboratory, usually from its own results | Set by the provider, from a reference value or a consensus |
| How often | Every run, or at set intervals | A few times a year |
| Sees | Imprecision, a sudden shift, a slow drift | Bias, method differences, clerical and interpretation errors |
| Misses | A bias present when the targets were set | A short-lived fault between distributions |
Internal quality control compares today with the laboratory’s own yesterday. If yesterday was already wrong, it has nothing to object to.
Example: in control all month, and biased
A laboratory changes calibrator lot and resets its control targets from its own new results. Internal quality control stays in control all month, because the targets moved with the calibration. The next EQA distribution returns z = +2.6: the outside value shows the shift that the laboratory’s own targets had absorbed.
EQA cannot release a run: it samples a few days a year, so it catches a short-lived fault late or not at all. Neither tool replaces the other.
How an EQA round runs
- Enrol in the scheme for the examinations you perform, and tell the provider your method, instrument and reagent for each.
- Receive the samples and check them on arrival: packaging, temperature, labels and the instructions. Report any problem to the provider before testing.
- Store and prepare them exactly as the instructions say. A wrong reconstitution volume shifts every result.
- Test them like patient samples: the staff who normally do the work, the routine method, and the routine number of measurements.
- Submit results by the closing date, in the units asked for, with the right method codes.
- The provider evaluates. It sets the assigned value, scores every result and compiles the statistics for each method group.
- Read the report in a fixed order and record the review.
- Act on anything unusual, and keep the record.
How results are judged
Two choices by the provider sit behind every verdict: the assigned value, and the limit a result is allowed to differ from it.
- The assigned value can come from a reference measurement, from how the sample was made, or from the consensus of the participants’ own results, either all methods together or within each method group. See how the assigned value is set.
- A z-score divides the difference by a standard deviation the scheme has chosen. A size of 2.0 or less is usually satisfactory, between 2.0 and 3.0 questionable, and 3.0 or more unsatisfactory. Some reports call a similar figure, based on the peer group’s own spread, a standard deviation index. See z-scores in proficiency testing and EQA.
- A percent difference against a fixed limit is common in clinical schemes. The limit is set before the round from what the result is used for, such as a clinical performance specification, and a result is either inside it or not.
- Qualitative results, such as detected or not detected, a blood group or an identification, are compared with an expected answer. Interpretive exercises are judged by an expert panel against criteria agreed in advance, and some are educational and not scored.
Peer groups and why the sample matters
EQA samples are often stabilised, freeze-dried, pooled or spiked so that they survive the post and last the round. That processing can make a sample behave differently from a fresh patient sample on some methods. A sample that different methods respond to in the same way as they do to patient samples is called commutable.
When a sample is not commutable, method groups can disagree on it even though they agree on patients. Scoring every method against one value would then blame laboratories for a property of the material, so many schemes score each laboratory against its own peer group: laboratories using the same method principle, instrument or reagent.
- Close to your group, and the group close to the target: no sign of a problem.
- Close to your group, but the group is off: a method bias shared by the group. Your users still receive it, so find out whether it is real on patient samples.
- Far from your group: look in your own laboratory.
Agreeing with a peer group shows consistency with peers, not trueness. Only a reference value on a commutable sample shows which group is right. How to read a PT or EQA report works through an example.
What ISO 15189 asks
ISO 15189:2022 deals with EQA in clause 7.3.7.3. In plain words, it expects a medical laboratory to:
- take part in EQA programmes suited to its examinations and to the way it interprets results;
- handle EQA samples as far as possible like patient samples, tested by the staff who normally do the work;
- not send EQA samples to another laboratory for confirmation before the results are submitted;
- use an alternative approach, with objective evidence, when no suitable programme exists;
- review its EQA results with the staff concerned, and act when the criteria are not met.
Taking part fairly also means not comparing or sharing results with other participants until the provider has issued its report. The pillar guide sets out the rules under taking part fairly.
Note: check the current edition and your accreditation body
Clause numbers change between editions, and accreditation bodies and regulators may add rules of their own, such as which schemes to use or how often. Read their current policy alongside this guide.
Choosing an EQA scheme
- Accreditation. Look for a provider whose ISO/IEC 17043 accreditation includes that scheme in its scope.
- The right examinations and levels. The scheme should cover what you report, at concentrations near your clinical decision points, not only in the comfortable middle of the range.
- Frequency. Enough distributions a year to catch a problem in reasonable time, and enough samples in each to tell a constant error from a proportional one.
- How the target is set. A reference value on a commutable sample says more than a consensus. Where a consensus is used, ask how many results it rests on.
- Your peer group. A group of three or four laboratories gives an unstable comparison.
- The report. It should state the assigned value and how it was set, the limit or standard deviation used, your score, and the statistics for your method group.
- Practicalities. Sample stability in transit to your location, the closing dates, and how results are submitted.
To see which providers offer a scheme for a given field or country, search the PT directory. It lists clinical EQA providers alongside PT providers in other fields.
When no EQA scheme exists
For a rare examination there may be no scheme, or none that fits. Record the search, then choose an alternative that compares the laboratory with something outside its own system, with the criteria fixed before the comparison:
- an exchange of samples with one or a few other laboratories;
- split patient samples tested by another laboratory, ideally one using a different method;
- a reference material with an assigned value, tested as an unknown;
- material tested earlier, compared with its previous value, if it is still stable.
Record and review the results the same way as EQA results, and search for a scheme again each year.
Acting on EQA results
- Every report: check that the provider scored what you meant to send, and record the review. See how to read a PT or EQA report.
- Across distributions: chart your scores. A run on one side of zero is a bias even when every score is satisfactory. See PT trends and Youden plots.
- An unsatisfactory result: treat it as a nonconformity, find the root cause and consider the patient results already reported. See investigating an unsatisfactory PT result.
For organisers of EQA schemes
Hospital groups, professional societies and reference laboratories often organise EQA themselves, for a network of sites or for a specialty no commercial scheme covers. The provider’s side has its own standard, ISO/IEC 17043, and its own statistics, described in ISO 13528. Running a proficiency testing scheme covers the scheme plan, the round and the report, and homogeneity and stability testing covers the checks on the samples before they are sent.
Take the free lesson
Module 1 · Proficiency testing, explained
Four short lessons and a practice case, with a check after each: what PT and EQA are, how they differ from internal QC, participation requirements and the provider’s role, and fair participation. The course is free in the EqasOnline Academy; lessons open after you create a free account.
Take the free lessonRun an EQA scheme?
EqasOnline lets PT and EQA providers run their schemes end to end (rounds, samples, results, ISO 13528 statistics, reports and certificates) in one system, with the evidence kept as the round runs. Labs enter results in their own portal, peer groups are formed by method, instrument and reagent, and each lab sees its own result against anonymised peers only after you release.