Running a proficiency testing scheme: the plan, the round and the report

A proficiency testing round involves many decisions: what to send, to whom, how to check it, how to score it and when to report. ISO/IEC 17043 expects a provider to make and document them before the round runs. This guide walks through the scheme plan, the round itself and the participant report, in the order a provider meets them.

Byline
By the EqasOnline team
Last reviewed
Reading time
10 min read
On this page
  1. Start with the scheme plan
  2. Choosing the levels and the matrix
  3. Fix the evaluation before the results arrive
  4. A round from plan to report
  5. Instructions to participants
  6. Closing dates and late results
  7. Confidentiality and participant codes
  8. Deterring and detecting collusion
  9. What the participant report contains
  10. After the report

Start with the scheme plan

The document that records a provider’s decisions is the scheme plan; some providers call it the scheme protocol or design. It has two readers. Inside the provider, it tells the people who prepare items, run the statistics and sign the report what to do. Outside, it is the rulebook: a participant can see in advance how its results will be judged, and an assessor can check afterwards that the provider kept to it.

What a scheme plan settles, with an invented example for each topic.
TopicWhat the plan settlesExample
Purpose and participantsWhat the scheme is for, who may enrol, how many are expectedMycotoxins in cereals for food testing laboratories; about 60 participants
MeasurandsWhat is measured, the unit and the decimals to reportDeoxynivalenol in µg/kg, whole numbers
Items and levelsMatrix, number of levels, the range coveredTwo naturally contaminated wheat flours per round
Homogeneity and stabilityWhich checks, which method, which criteriaThe ISO 13528 Annex B checks
Assigned valueHow xpt is set and how its uncertainty is estimatedRobust mean of participants’ results
σptHow it is setA fixed percentage of xpt, chosen from what the results are needed for
ScoringWhich score, which limits, the rounding rulez, with a size of 2.0 or less satisfactory
TimetableDispatch, closing date, report dateClosing date 21 days after dispatch
ReportingWhat each participant receives, and whenAn individual report within 21 days of the close
ConfidentialityCodes, who sees identities, what is publishedRandom codes; the key held by named staff
ContingenciesLost or damaged items, too few results, failed checksFewer than 12 results: report for information, no scores
What a scheme plan settles, with an invented example for each topic.

Three of those rows have guides of their own: homogeneity and stability testing, the assigned value and choosing σpt.

Choosing the levels and the matrix

A level is the concentration, or value, of the measurand in an item. Good choices follow the laboratories’ real work.

  • Cover the range in use, not one convenient concentration.
  • Sit near decision points: a legal limit, a clinical cut-off, a product specification. An error matters most where it changes a decision.
  • Include the low end if laboratories report results near their limit of quantification, and a negative or blank item where false positives matter.
  • Send two or more levels so that constant and proportional errors can be told apart later.

Example: HbA1c

Laboratories use 48 mmol/mol as a diagnostic cut-off and see patients from about 30 to 120 mmol/mol. A scheme that always sends one level near 95 mmol/mol says little about performance at 48, where a small bias changes a diagnosis. Levels near 40, 50 and 80 mmol/mol across the year cover the cut-off and the range.

The item matrix should resemble routine samples: a real soil rather than spiked sand, a milk powder rather than a solution of the analyte, serum rather than buffer. A matrix that is too clean skips steps the laboratory normally performs, such as extraction, digestion or clean-up, so the round tests less than it seems to.

In clinical schemes it also matters whether the item is commutable: whether different methods respond to it the way they respond to patient samples. Freeze-drying, stabilisers and spiking can break this. On an item that is not commutable, method groups can disagree even though they agree on patients, so scoring every method against one reference value would blame laboratories for a property of the material. The plan says whether this has been checked and, if not, whether results are scored within method groups.

Fix the evaluation before the results arrive

The plan fixes how xpt and σpt will be set, which score and limits apply, how outliers are handled, and what happens when something goes wrong. The numbers, such as this round’s robust mean, are calculated after the close, but the method that produces them does not change once results are seen. If the method could be chosen afterwards, almost any outcome could be produced, and scores would mean something different in every round.

Example: a yardstick changed too late

A wastewater scheme fixes σpt for ammonium at 0.15 mg/L. In one round the results spread widely, with a robust standard deviation of 0.32 mg/L. A laboratory 0.50 mg/L above xpt scores 0.50 / 0.15 = +3.3, unsatisfactory. Rescored with 0.32 mg/L it would be 0.50 / 0.32 = +1.6, satisfactory. The laboratory did nothing differently; only the yardstick moved.

If a plan proves unsuitable, the provider changes it for future rounds, tells participants before those rounds, and records why. Within the current round, anything unexpected is handled by the contingency rules already in the plan and stated in the report.

A round from plan to report

An example timetable for one round, counted in weeks from dispatch.
WhenStep
Week −16Plan approved; participants enrol
Week −10Items prepared and packaged
Weeks −6 to −2Homogeneity check on packaged items; dispatch only after it passes
Week 0Dispatch with instructions and the closing date
Weeks 0 to 3Participants test the items as routine samples
Week 3Closing date; stability items measured
Weeks 4 to 5Evaluation by the planned method
Week 6Reports issued; laboratories review their results
An example timetable for one round, counted in weeks from dispatch.

Instructions to participants

The provider gives participants detailed instructions with the items. Good instructions answer every practical question before anyone has to ask it.

What participant instructions cover, and why.
TopicWhat to sayWhy
On receiptCheck the packaging and item, record the arrival date and temperature, and contact the provider before testing if anything is wrongA damaged item can be replaced instead of scored
StorageTemperature and the longest time before testingStability was shown only under these conditions
PreparationReconstitution volume and diluent, standing time, mixingA wrong volume shifts every result
TestingTreat the item as a routine sampleThe round should test routine work
ReportingUnit, decimals, method or instrument code, how to report results under a reporting limitResults must be comparable and correctly grouped
Closing dateDate, time and time zone, and how to submitOne cut-off for everyone
SafetyHazards of the materialStaff handle it safely
What participant instructions cover, and why.

Example: one missing phrase

A drinking-water scheme asks for nitrate “in mg/L” without saying whether as nitrate or as nitrogen. The same item comes back as 44.3 from some laboratories and 10.0 from others, a factor of 4.43 (62.0 / 14.0). Neither group made an analytical error, yet whichever group is smaller looks unsatisfactory against the consensus. One phrase in the instructions, “as NO₃⁻”, would have prevented it.

Closing dates and late results

The scheme plan fixes the closing date and says what happens to late results. Common rules are that late results are not scored, or that they are evaluated for information only, clearly marked, and never used to calculate xpt.

At the close, the data set is frozen. The provider analyses the results it holds, using the method in the plan, and nothing changes afterwards: no late additions, no corrections to submitted results unless the plan allows for them, no re-running with a different data set. Once xpt or the report is out, a late or corrected result could simply be tuned to it.

The same rule applies to everyone. If the provider knows of a delivery problem before the close, it can extend the closing date for every participant affected, announced in advance. A laboratory that finds its own error after the close keeps the submitted result in the report, and records and investigates the error itself.

Confidentiality and participant codes

  • Each participant is known by a code. The key that links codes to laboratories is held by the few staff who need it to run the scheme.
  • Codes are allocated at random, not in order of enrolment. Some schemes issue a new code each round; where codes are permanent, the key needs tighter control.
  • Never published: a laboratory’s name next to its results, or anything that identifies it indirectly, such as a method group of one.
  • Disclosure to a regulator or accreditation body happens only with the participant’s agreement, where the law requires it, or where the scheme’s terms, made known to participants in advance, provide for it. The participant is told what was released.

Deterring and detecting collusion

Collusion is participants comparing or sharing results before the close. Falsification is reporting a result that was not obtained by testing the item as a routine sample: copied, invented, or nudged towards an expected value. Either way the score stops saying anything about the laboratory, and copied results can distort a consensus xpt for everyone else. Providers are expected to take measures against both.

  • Different items or levels for participants likely to be in contact: the same city, the same hospital group, the same owner. Allocate at random and keep the allocation confidential.
  • Release nothing before the close: no assigned values, expected ranges, summary statistics or early reports.
  • Random codes, so participants cannot find one another in a report.
  • Do not reuse a material whose value has been published, unless enough time has passed and nothing links it to the earlier round.
  • Clear terms that say what happens when collusion or falsification is found.

No single pattern proves anything, but some deserve a closer look: identical results across several measurands, especially to more decimals than the method supports; identical errors, such as the same unit slip or the same swapped digits; a result that matches a level the participant did not receive; and two participants whose results track each other, round after round, far more closely than anyone else’s.

Example: the wrong level

In a round for copper in sediment, level A (xpt = 48.6 mg/kg) and level B (xpt = 31.2 mg/kg) are allocated at random, with σpt = 2.5 mg/kg for level B. One participant received level B but reports 48.9 mg/kg: z = (48.9 − 31.2) / 2.5 = +7.1 against its own item, yet within 0.3 mg/kg of level A, which went to another participant in the same city. The provider first checks its own dispatch and labelling records, because a label mix-up would be its own error, then follows its documented procedure for suspected collusion.

What the participant report contains

ISO/IEC 17043 lists what a PT report includes, unless the provider documents a reason to leave something out. Paraphrased and grouped:

The contents of a participant report, grouped.
GroupContents
IdentityProvider and contact details; scheme and round; report number, issue date and status, for example final or amended; page numbers and a clear end
The participantIts code, the items it received, and its results as reported
The itemsWhat they were and how they were prepared; the homogeneity and stability outcomes
The evaluationxpt, how it was set, its uncertainty and its traceability; σpt and how it was set; the scores and their criteria; the number of participants and results used; summary statistics and graphics
InterpretationComments on performance and advice on reading the statistics
AuthorisationName and role of the person who authorised the report; any subcontracted activities; a confidentiality statement
The contents of a participant report, grouped.

A complete report lets a laboratory recalculate its own score, judge how reliable xpt is, and show an assessor exactly what was evaluated. How to read a PT report is the participant’s side of the same list.

After the report

A round does not end when the report is issued. Participants may question an evaluation, and a provider is expected to have a documented process for receiving, investigating and deciding on complaints and appeals, to keep the participant informed, and to tell it the outcome. If an evaluation was wrong, the provider issues a corrected report, marked as amended, and keeps the earlier version on record.

What remains is the record: who enrolled, what was sent and when, whether the items passed their checks, which results arrived, how they were scored, and which report version each participant received. Most of the work in a round is keeping that record as the round runs, not rebuilding it before an assessment. For the wider picture of what PT is and what the standards ask, see proficiency testing: the complete guide.

Take the free lesson

Module 2 · Inside a PT round

Five lessons and a practice case: the scheme plan, homogeneity with an interactive calculator, stability and transport, running the round, and collusion and the participant report. The course is free in the EqasOnline Academy; lessons open after you create a free account.

Take the free lesson

Run a PT scheme?

EqasOnline lets PT and EQA providers run their schemes end to end (rounds, test items, results, ISO 13528 statistics, reports and certificates) in one system, with the evidence kept as the round runs. Rounds sit on one calendar with deadlines in your time zone, reminders go out three days and one day before the deadline, participants appear as codes in your screens, and reports go out on your release date.