Reports and follow-up

How to read a PT or EQA report

A PT report holds more than a row of scores. Read in a fixed order, it tells you whether the provider scored the result you meant to send, how much weight each score can bear, and whether a difference belongs to your laboratory or to your method. This guide gives the checklist, works through an example report, and shows how to read peer-group statistics.

Byline
By the EqasOnline team
Last reviewed
Reading time
7 min read
On this page
  1. Read it in the same order every time
  2. The checklist
  3. Back-calculate the score
  4. How much weight a score can bear
  5. A worked extract
  6. Reading the peer-group statistics
  7. When your group is off, and when you are
  8. What group statistics cannot show
  9. After the review

Read it in the same order every time

Jump straight to the scores and you may spend a week investigating a method when the cause was a slipped digit. Read top to bottom, the same way every time, and the cheap explanations surface first.

Reviewing the report is also a requirement. ISO/IEC 17025:2017 (7.7.3) and ISO 15189:2022 (7.3.7.3) both expect a laboratory to review its PT or EQA results and act on them.

The checklist

Eight checks, in order, and what each one catches.
StepWhat to checkWhat can go wrong
1. IdentityParticipant code, scheme, round, report versionSomeone else’s report; a version that has since been amended
2. ItemsItem numbers against your receipt recordResults entered against the wrong item
3. Your entriesResult, unit, decimals and method code against your worksheetA slipped digit, the wrong unit, the wrong method group
4. The scoreBack-calculate the scored resultThe value scored is not the value you measured
5. The assigned valueHow it was set, and from how many resultsA consensus resting on few results
6. Its uncertaintyCompare it with 0.3 σptAn uncertain assigned value scored as if it were exact
7. σptWhere it came fromA spread taken from this round, not a fixed requirement
8. CommentsHomogeneity, stability, exclusions, late resultsA score reported for information only
Eight checks, in order, and what each one catches.

Back-calculate the score

For a z-score, the result that was scored is:

x = xpt + z × σpt

For a z′ score the denominator includes the uncertainty of the assigned value, so the multiplier is √(σpt2 + u2(xpt)) instead of σpt. If the recovered value matches your entry, the provider scored what you sent, and any gap from your worksheet lies in your own trail. If not, ask the provider. Expect agreement only within rounding: a z given to one decimal fixes x to about ±0.05 σpt.

How much weight a score can bear

  • How the assigned value was set. A value from formulation or a reference measurement does not depend on the participants. A consensus does, and a robust mean of 64 results is far firmer than one of 12. See how the assigned value is set.
  • Its uncertainty. When u(xpt) is no more than 0.3 σpt it is negligible. Above that, the scheme should use z′, flag the score, or not score it.
  • Where σpt came from. A σpt fixed in the scheme plan means a z of +1.5 says the same thing every round. A σpt taken from this round’s robust standard deviation only places you among this round’s participants: roughly 95 % of them will look satisfactory whatever the general standard of the round. See choosing σpt.

A worked extract

Drinking-water chemistry, one round, report version 1. Every σpt was set in the scheme plan.

An extract from a participant report. Example data.
MeasurandResult as enteredxpt (how set)u(xpt)σptScore
Nitrate, mg/L38.437.6 (robust mean of 64 results)0.251.6z = +0.5
Fluoride, mg/L0.910.85 (formulation)0.0050.05z = +1.2
Aluminium, mg/L0.150.100 (formulation)0.0020.010z = +5.0
Manganese, µg/L52.748.0 (robust mean of 12 results)2.04.8z′ = +0.9
Sulfate, mg/L131124 (robust mean of 58 results)1.26.2z = +1.1
An extract from a participant report. Example data.

The report’s comments say that manganese is scored with z′ because u(xpt) exceeds 0.3 σpt, and that homogeneity and stability met the scheme’s criteria for every measurand. The laboratory’s worksheet reads 38.4, 0.91, 0.105, 52.7 and 131.

  • Entries. Four lines match the worksheet. Aluminium does not: 0.105 was measured and 0.15 was entered, a dropped zero.
  • Back-calculation. 5.0 × 0.010 above 0.100 is 0.150 mg/L, exactly the entry, so the provider scored what was sent. The worksheet value would have scored (0.105 − 0.100) / 0.010 = +0.5.
  • The z′ line. √(4.8² + 2.0²) = √27.04 = 5.2, and 0.9 × 5.2 above 48.0 is 52.68, which rounds to the 52.7 µg/L entered. Using σpt alone gives 52.32, a mismatch that does not exist.
  • The values behind the scores. For nitrate, u(xpt) = 0.25 is under 0.3 × 1.6 = 0.48. For manganese, 2.0 is above 0.3 × 4.8 = 1.44 and the consensus rests on 12 results, so that score deserves more caution.

The aluminium score stands: results cannot be changed after the closing date. Record it as a nonconformity with a clerical cause and find out why the slip got through. The worksheet suggests the measurement was fine, but this round’s score is no evidence either way.

Warning: check every line, not only the flagged ones

A slip does not always produce a bad score. Had 0.115 been typed, the line would show z = +1.5 and nobody would look twice.

Reading the peer-group statistics

Many reports summarise results by group, built from the method codes participants enter: the method principle, the instrument or platform, and the reagent, kit or calibrator. Each group usually gets a count, a median or robust mean, and a robust standard deviation. Your score may be against the all-method assigned value, a reference value or your group’s own value; the scheme plan says which, and your method code decides your group.

The useful step is to split your difference into two parts:

x − xpt = (x − group median) + (group median − xpt)

The first part is yours alone. The second is shared by everyone using your method. Divide each by σpt to read them on the z scale.

Example: ammonium in wastewater

xpt = 2.40 mg/L (formulation) and σpt = 0.12 mg/L. The phenate group’s median is 2.38 mg/L from 32 results, and the ion-selective electrode group’s is 2.71 mg/L from 10 results.

Two questionable scores with different causes. Example data.
LaboratoryResultzYou − groupGroup − xpt
P (electrode)2.74+2.83+0.03 (0.25 σpt)+0.31 (2.58 σpt)
Q (phenate)2.70+2.50+0.32 (2.67 σpt)−0.02 (0.17 σpt)
Two questionable scores with different causes. Example data.

Both are questionable, with almost the same score. P agrees with its group and the group reads high: a method bias. Q is far from a group that reads correctly: look in Q’s laboratory.

Reading the two parts of the difference.
You against your groupYour group against xptReading
CloseCloseNo sign of a problem
CloseOffA method bias, shared by the group
OffCloseLook in your own laboratory
OffOffBoth: deal with each part separately
Reading the two parts of the difference.

When your group is off, and when you are

A method bias is not your laboratory’s fault, but your users still receive it. Ask whether it is known, from manufacturer notices, earlier rounds or published comparisons; whether it matters for a clinical cut-off, a regulatory limit or a specification; and whether it is real on routine samples, because a material that is not commutable can create a group difference that patients and customers never see. Then record it, tell users if it matters, and take it up with the supplier or consider another method.

Warning: never recalibrate to a PT value

One round’s assigned value is not a calibrator, and a nudge towards it hides the problem. If your scheme scores you against your group’s value, a satisfactory score shows consistency with peers, not trueness.

When you are off from your group, your peers on the same method got close to the group median, so the method itself is not the explanation. Look inside the laboratory: the clerical trail, calibration, reagent and calibrator lots, internal quality control, and how the item was handled.

What group statistics cannot show

  • Which group is right, unless xpt comes from a reference value. An all-method consensus leans towards the largest group: if 80 % of results come from a method reading 5 % high, the plain mean sits about 4 % high, and correct methods look low.
  • Anything firm from a tiny group. The standard uncertainty of a group median is roughly 1.25 × s / √n. With four results and s equal to σpt, that is about 0.63 σpt, so a median 1 σpt from xpt proves little, and your own result is a quarter of the group.
  • The right comparison if your method code is wrong. You are then measured against the wrong peers.
  • Differences hidden inside one label. Calibrators, lots and settings can vary within one method name.

After the review

Record what you checked and what you concluded, even when every score is satisfactory. Then place the scores on a chart across rounds, where a small, steady problem shows itself long before it produces a failed score: see PT trends and Youden plots. For an unsatisfactory score, or a pattern, see investigating an unsatisfactory PT result.

Take the free lesson

Module 5 · Reading reports and trends

Four lessons and a practice case: reading a PT report, peer groups and method bias, trends across rounds, and Youden plots with an interactive tool. The case follows three years of creatinine results through a reagent change. The course is free in the EqasOnline Academy; lessons open after you create a free account.

Take the free lesson

Issuing PT reports?

In EqasOnline reports go out on your release date, under your logo, seal, signatures and document control code, and one click notifies every lab. Each lab sees its own result against anonymised peers, with peer groups by method, instrument and reagent, peer histograms, box plots and a chart of its z-scores across rounds.